Power Consumption

When installed and connected to appropriate electrical power sources, the Luna Network HSM 7 draws power as follows:

Activity Draw
Standby (connected to AC electrical mains but not powered on)
26W (typical)
Power-on inrush surge

15A (typical)

40A at 90-132VAC (max)

60A at 180-265VAC (max)

Active (under load from clients)

100W (idle)

105W (max)

The Luna Network HSM 7 appliance has two power supplies, each rated at 300W, either of which is capable of running the system alone.